http://arxiv.org/abs/1411.1118
The sensitivity of a THz hot-electron nanobolometer (nano-HEB) made from a normal metal is analyzed. Johnson Noise Thermometry (JNT) is employed as a readout technique. In contrast to its superconducting TES counterpart, the normal-metal nano-HEB can operate at any cryogenic temperature depending on the required radiation background limited Noise Equivalent Power (NEP). It does not require bias lines; 100s of nano-HEBs can be read by a single low-noise X-band amplifier via a filter bank channelizer. The modeling predicts that even with the sensitivity penalty due to the amplifier noise, an NEP ~ 10$^{-20}$ – 10$^{-19}$ W/Hz$^{1/2}$ can be expected at 50-100 mK in 10-20 nm thin titanium (Ti) normal metal HEBs with niobium (Nb) contacts. This NEP is fairly constant over a range of readout frequencies ~ 10 GHz. Although materials with weaker electron-phonon coupling (bismuth, graphene) do not improve the minimum achievable NEP, they can be considered if a larger than 10 GHz readout bandwidth is required.
B. Karasik, C. McKitterick, T. Reck, et. al.
Thu, 6 Nov 14
14/67
Comments: 5 pages, 5 figures, presented at the 25th International Symposium on Space Terahertz Technology (Moscow, Russia, 27-30 April 2014)
You must be logged in to post a comment.